PERFORMANCE PROFILE
Flow Range
|
• Constant flow from 1000 to 5000 ml/min (5 to 500 ml/min requires low flow holder) |
Compensation Range (Back Pressure Capability)
|
• 5000 ml/min at 20 inches water back pressure • 4000 ml/min at 30 inches water back pressure • 3000 ml/min at 40 inches water back pressure • 2000 ml/min at 50 inches water back pressure • 1000 ml/min at 60 inches water back pressure |
Power
|
• Removable rechargeable lithium-ion (Li-Ion) battery, 7.4 V, 2.6 Ah, 19.2 Wh or AC using cradle |
Run Time
|
• 40+ hours at 2000 ml/min* • 15+ hours at 5000 ml/ min* • Indefinite run from charging cradle |
Accuracy
|
• Flow control: ± 5% of set-point after calibration to desired flow
• Atmospheric pressure: ± 0.3 in Hg • Temperature: ± 1.0 C |
Display/Parameters
|
• High-contrast backlit LCD/Time, date, battery status, flow rate, sample volume, temperature, atmospheric pressure, back pressure, programmed run remaining time, and elapsed run time |
User Interface
|
• Eight-area capacitive touch screen with auto-dim and locking options |
Sound Level
|
• Average 51.7 dB at 3-ft (1-m) distance using a 37-mm, 0.8-µm MCE filter cassette |
Tubing
|
• Requires ¼-inch ID tubing |
Dimensions/Weight
|
• 4.1 x 3.7 x 2.8 in (10.4 x 9.4 x 7.1 cm),19.4 oz (550 gm) |
Certifications
|
• Intrinsic safety (SKC Cat. No. 220-4000 operated with SKC Battery Pack P75718)
• Class I, Groups A, B, C, D; Class II, Groups E, F, G; Class III, T4; Class I, Zone 0 • AExia IIC T4 Ga; Exia IIC T4 Ga; -20°C ≤ Ta ≤ 45°C; Ex II 1G Exia IIC T4 Ga; IECEx UL 19.0100 • DEMKO 19ATEX 2288 • CE 0539 • RoHS compliant • Designed to meet ISO 13137:2013 |