AirChek XR5000

Discover the perfect combination of features and power in this compact, lightweight pump that gets the job done. The AirChek® XR5000 air sample pump with two Li-Ion battery options provides powerful flows for particulate sampling, low flows for gas/vapor sampling, and the long run-times needed for sampling low-concentration hazards.

Performance Profile

Flow Range
1000 to 5000 ml/min  (5 to 500 ml/min requires optional low flow kit)
Compensation Range
5000 ml/min at 10 inches water back pressure
4000 ml/min at 20 inches water back pressure
2000 ml/min at 50 inches water back pressure

Typical Back Pressure of Sampling Media (inches water)

Filter/Pore Size (µm)
Flow Rate (L/min)
1.
1.5
2
2.5
3
3.5
4
5
25-mm MCE/0.8
6
9
12
15
18
21
25
31
25-mm MCE/0.45
14
22
28
35
40
44
50
63
37-mm MCE/0.8
2
3
4
5
6
7
9
11
37-mm PVC/5.0
1
1
2
2
2.5
3
3
4
37-mm polycarbonate/0.45
4
6
8
10
12
15
17
21
25-mm MCE/0.45 microvacuum
21
31
40
48
59
69
79
100
37-mm PTFE/1.0
1.5
2.5
4
5.5
7
8
9.5
12

Compare the information in this table to pump compensation range to determine appropriate applications.

Flow Compensation System
Isothermal closed loop flow sensor
Tubing
Requires 1/4-inch ID tubing
Accuracies
Timing: 1 min/mo at 25 C
Flow Rate: ± 5% of set-point after calibration to desired flow
Battery Charge Level Indicator
Icon displays on LCD at full, mid, low charge, imminent low battery fault, and low battery fault.
Temperature Range
Operating: 32 to 104 F (0 to 40 C)
Charging: 32 to 113 F (0 to 45 C)
Storage: -4 to 95 F (-20 to 35 C)
Typical Run Times
XR5000 Model
2 L/min
5 L/min
High-power Li-Ion
40 hrs
22 hrs
Standard Li-Ion
20 hrs
11 hrs
Timer Display Range
1 to 9999 minutes (6.8 days). If run time exceeds 6.8 days, timer display rolls over.
Flow Fault
If pump is unable to compensate for > 15 seconds due to excessive back pressure, the pump stops and holds run time display. Auto-restart is attempted every 15 seconds up to 5 times.
Low Battery Fault
15 sec to sleep
Auto-off
5 min of inactivity
Battery Pack
(model dependent)
High-power Li-Ion (4 cell), rechargeable, 7.4 V, 5.2 Ah, 38.5 Wh capacity
or
Standard Li-Ion (2 cell), rechargeable, 7.4 V, 2.6 Ah, 19.2 Wh capacity
Battery Recharge Time
(with SKC-approved charger)
Standard Li-Ion (2 cell): approximately 4 hrs
High-power Li-Ion (4 cell): approximately 8 hrs
Size
High-power Li-Ion model: 5.5 x 3 x 2.3 in (14 x 7.6 x 5.8 cm)
Standard Li-Ion model: 4.3 x 3 x 2.3 in (10.9 x 7.6 x 5.8 cm)
Weight
High-power Li-Ion model: 21 oz (0.6 kg)
Standard Li-Ion model: 16 oz (0.45 kg)
Case
Anti-static plastic
RFI/EMI Shielding