| Main Unit Specifications | |
|---|---|
| Measurement Units | PPM or mg/m3 |
| Sensor Head | Removable and Replaceable |
| Output | 0 to 5 V, Alarm output |
| Data Logging | 8188 data points |
| Power Supply | Li-Ion Battery or AC Adapter |
| Dimensions (with sensor head) | 7.0 x 4.75 x 2.0 inches (195 x 122 x 54 mm) |
| Weight (with sensor and battery) | 1 lbs (460 g) |
| Standards | CE-mark Part 15 of FCC Rules EN 50082-1: 1997 EN 50081-1: 1992 |
| Warranty | 1 year |
| What’s Included | Meter, Li-ion battery pack, Operation manual, AC adapter, PC connection cable*, Data processing software*
(*PC connection cable and data processing software only available for Model 2750) |
| Gas Monitor Sensor Heads | ||||||
|---|---|---|---|---|---|---|
| Sensor | Range (PPM) | Accuracy of Calibration | Min. Detection Limit (PPM) | Resolution (PPM) | Operational Temperature Range | Operational RH Range |
| Ammonia | 0-100 | <±10% | 0.2 | 0.1 | 32-104°F | 10-90% |
| Ammonia (leak) | 0-1000 | <±15% | 2 | 1 | 32-104°F | 10-90% |
| Carbon Monoxide | 0-25 | <±0.5 ppm 0-5 ppm; | 0.05 | 0.01 | 32-104°F | 10-90% |
| <±10% 5-25 ppm | ||||||
| Carbon Monoxide | 0-100 | <±1 ppm 0-10 ppm; | 0.2 | 0.1 | 32-104°F | 10-90% |
| <±10% 10-100 ppm | ||||||
| Carbon Monoxide | 0-1000 | <±10 | 1 | 1 | 32-104°F | 10-90% |
| Carbon Dioxide | 0-2000 | <±10 ppm +5% | 10 | 1 | 32-104°F | 10-90% |
| Carbon Dioxide | 0-5000 | <±20 ppm +5% | 20 | 1 | 32-104°F | 10-90% |
| Chlorine | 0-10 | <±10% | 0.02 | 0.01 | 32-104°F | 10-90% |
| Formaldehyde | 0-10 | <±0.05 ppm 0-0.5 ppm; | 0.01 | 0.01 | 32-104°F | 10-90% |
| <±10% 0.5-10 ppm | ||||||
| Hydrogen | 0-5000 | <±10 | 5 | 1 | 32-104°F | 10-90% |
| Hydrogen Sulfide | 0-10 | <±0.05 ppm 0-0.5 ppm; | 0.04 | 0.01 | 32-104°F | 10-90% |
| <±10% 0.5-10 ppm | ||||||
| Hydrogen Sulfide | 0-100 | <±0.5 ppm 0-5 ppm; | 0.04 | 0.01 | 32-104°F | 10-90% |
| <±10% 5-100 ppm | ||||||
| Methane | 0-10000 | <±15% | 10 | 1 | 32-104°F | 10-90% |
| Nitrogen Dioxide | 0-1 | <±0.02 ppm 0-0.2 ppm; | 0.005 | 0.001 | 32-104°F | 10-90% |
| <±10% 0.2-1 ppm | ||||||
| Ozone | 0-0.15 | <±0.005 ppm | 0.001 | 0.001 | 32-104°F | 10-90% |
| Ozone | 0-0.5 | <±0.008 ppm 0-0.1 ppm; | 0.001 | 0.001 | 32-104°F | 10-90% |
| <±10% 0.1-0.5 ppm | ||||||
| Ozone | 0-10 | <±0.1 ppm 0-1 ppm; | 0.02 | 0.01 | 32-104°F | 10-90% |
| 15% 1-10 ppm | ||||||
| Perchloroethylene | 0-200 | <±5 ppm 0-50 ppm; | 1 | 1 | 32-104°F | 10-90% |
| <±10% 50-200 ppm | ||||||
| Sulphur Dioxide | 0-10 | <±0.05 ppm 0-0.5 ppm; | 0.04 | 0.01 | 32-104°F | 10-90% |
| <±10% 0.5-10 ppm | ||||||
| Sulphur Dioxide | 0-100 | <±0.5 ppm 0-5 ppm; | 0.4 | 0.1 | 32-104°F | 10-90% |
| <±10% 5-100 ppm | ||||||
| VOC (GSS sensor) | 0-25 | <±10% | 0.1 | 0.1 | 32-104°F | 10-90% |
| VOC (GSS sensor) | 0-500 | <±10% | 1 | 1 | 32-104°F | 10-90% |
| VOC (GSS sensor) | 0-20 | <±10% | 0.02 | 0.01 | 32-104°F | 10-90% |
| VOC (GSS sensor) | 0-1000 | <±10% | 0.2 | 0.1 | 32-104°F | 10-90% |
| Temperature | -20-100°C | <± 0.3°C | 0.01°C | -40-248°F | 0-105% | |
| Humidity | 0 to 100% | <± 2% RH | 1% RH | -40-248°F | 0-105% | |
| Gas Monitor Laser Particle Sensor Head | ||||||
|---|---|---|---|---|---|---|
| Sensor Type | Range (mg/m3) | Accuracy of Calibration | Min. Detection Limit (mg/m3) | Resolution (mg/m3) | Operational Temperature Range | Operational RH Range |
| LPC | 0.001-1.000 | ±0.005 mg/m3 +15% | 0.001 | 0.001 | 0-40°C | 0-90% |